The ORTHO VISION
Intended Use
ORTHO VISION™ Analyzer automates in vitro Immunohematology
testing of human blood utilizing ORTHO BioVue® Cassette technology.
ORTHO VISION™ Analyzer:
• Automates test processing functions including liquid pipetting,
reagent handling, incubation, centrifugation, reaction grading
and interpretation and data management requirements using
ORTHO BioVue® Cassettes and digital image processing.
• Standalone instrument or interfaced to the Laboratory
Information System (LIS).
• Monitors critical processing monitoring and provides
for complete management of system maintenance and
quality control.
Installation and Site Specifications
Although trained service personnel install the ORTHO VISION™
Analyzer at the laboratory site, the site must be prepared according
to site specifications.
Physical Dimensions
ORTHO VISION™ Analyzer
Width: 107.4 cm (42.3 in.)
Depth: 77 cm (30.31 in.)
Height: 88.9 cm (35 in.)
Height with maintenance door open: 137 cm (54 in.)
ORTHO VISION™ Table Dimensions (highly recommended)
Table without shelf:
121.9 cm (48 in.) x 76.2 cm (30 in.) x 76.2 cm (30 in.)
Table with shelf at the front:
121.9 cm (48 in.) x 106.7 cm (42 in.) x 76.2 cm (30 in.)
- Occupies 61 cm (24 in.) on left side front of table width
Table with the shelf on the side:
152.4 cm (60 in.) x 76.2 cm (30 in.) x 76.2 cm (30 in.)
- Occupies 61 cm (24 in.) in depth on left side – side of table
Weights
ORTHO VISION™ Analyzer: 192 Kg (424 lbs)
ORTHO VISION™ Table: 120.2 kg (265 lbs)
Power Requirements One dedicated (3-wired single phase; line-neutral and single circuit) AC power line for connection to facility power. Input Voltage 100 - 240 V AC 50/60 Hz 1000 VA Environmental Specifications • Operating Temperature: 18 - 30° C (64.4 - 86.0° F) • Site Relative Humidity: 15 - 85% RH (non-condensing) • Maximum Altitude: 2438.0 m (8000 ft.)
• Heat Output: 3412 BTU/hr Sample and Reagent Load Door Liquid System Shelf Attachment Positions e-Connectivity® Remote Connectivity Provides the ability to connect your system to Ortho Clinical Diagnostics in a way that enables remote diagnostics as well as, monitor and review system configuration, data, and performance.
Power Requirements One dedicated (3-wired single phase; line-neutral and single circuit) AC power line for connection to facility power. Input Voltage 100 - 240 V AC 50/60 Hz 1000 VA Environmental Specifications • Operating Temperature: 18 - 30° C (64.4 - 86.0° F) • Site Relative Humidity: 15 - 85% RH (non-condensing) • Maximum Altitude: 2438.0 m (8000 ft.)
• Heat Output: 3412 BTU/hr Sample and Reagent Load Door Liquid System Shelf Attachment Positions e-Connectivity® Remote Connectivity Provides the ability to connect your system to Ortho Clinical Diagnostics in a way that enables remote diagnostics as well as, monitor and review system configuration, data, and performance.